To ensure the accuracy of the thermal imaging temperature measurement gate guard data and protect the equipment from damage, please read all the information below before using your device.
1. The optimal area for thermal imaging temperature detection is 2-3 meters, and it is recommended to install it at a height of about 1.5-2 meters.
2. Environmental changes have a significant impact on the temperature measurement accuracy of infrared thermography. It is recommended to choose a stable indoor environment and avoid deploying thermography thermometers in areas such as air vents, direct air conditioning, fans, or near heating equipment.
3. The background of the position targeted by the camera should avoid other high-temperature devices that are higher than human body temperature, such as heating air conditioners, water dispensers, high-power incandescent lamps, etc., to prevent continuous false alarms.
4. It is recommended to perform temperature calibration every day or every few days to maintain temperature accuracy without black body configuration.
5. When the thermal imager reaches thermal equilibrium 10-20 minutes after startup, temperature measurement is most accurate. Therefore, it is recommended to start temperature calibration and measurement 15 minutes after startup.
6. Place the temperature measurement area in such a way that one person appears at a time to prevent false alarms caused by multiple people entering at the same time.
7. Individual non patients with fever may trigger an alarm for excessive temperature, such as after drinking alcohol, engaging in vigorous exercise, drinking hot water, local facial inflammation, prolonged sun exposure, or coming out of a hot and humid environment. At this time, the test subject can be allowed to quiet down for a while or cool down before conducting measurement screening